Millivolts ↔ temperature with cold junction compensation, tolerances and IEC 60584-1 tables
The calculations use the official ITS-90 reference functions (IEC 60584-1, NIST), the same ones used to generate the printed tables: the result matches them over the whole range of each type.
Generates the table for the type selected above, with the reference junction at 0 °C. Each row adds (or subtracts, when negative) the digit of the column.
The curve of the selected type is highlighted. Base metal types (E, J, K, N, T) give much more signal than platinum types (R, S, B).
The resistance of the two conductors does not change the EMF if the instrument has a high input impedance, but it is worth knowing: a value much higher than expected reveals a loose joint or a damaged conductor, and old moving-coil indicators need it for calibration.
A thermocouple does not measure a temperature but the difference between the measuring junction and the point where its wires connect to the copper of the instrument (reference junction or cold junction). The tables are built with the reference junction at 0 °C, so the instrument must add the EMF that corresponds to the temperature of its terminals:
\(E(25\ °C) = 1.000\ \text{mV}\), so \(E(t) = 10.000 + 1.000 = 11.000\ \text{mV}\) and the table gives \(t = 270.7\ °C\). Without compensation you would read 246.2 °C: an error of almost 25 °C. You must add millivolts, not degrees: because the curve is not linear, adding 25 °C to the reading gives an approximate result, but not an exact one.
Transmitters and controllers do it automatically with a sensor (thermistor, Pt100 or semiconductor) next to the terminals. That is why the cable running to them must be extension or compensating cable of the same type: if it is joined to copper halfway, the cold junction moves to that joint, whose temperature nobody measures.
The tolerance is the maximum error allowed for a new thermocouple, with the reference junction at 0 °C. In the lower range the tolerance is fixed and in the upper range it is proportional to the temperature (\(|t|\) is the temperature in °C without sign).
| Type | Class 1 | Class 2 | Class 3 |
|---|---|---|---|
| T | −40 to 125 °C: ±0.5 °C 125 to 350 °C: ±0.004·|t| | −40 to 133 °C: ±1 °C 133 to 350 °C: ±0.0075·|t| | −67 to 40 °C: ±1 °C −200 to −67 °C: ±0.015·|t| |
| E | −40 to 375 °C: ±1.5 °C 375 to 800 °C: ±0.004·|t| | −40 to 333 °C: ±2.5 °C 333 to 900 °C: ±0.0075·|t| | −167 to 40 °C: ±2.5 °C −200 to −167 °C: ±0.015·|t| |
| J | −40 to 375 °C: ±1.5 °C 375 to 750 °C: ±0.004·|t| | −40 to 333 °C: ±2.5 °C 333 to 750 °C: ±0.0075·|t| | – |
| K and N | −40 to 375 °C: ±1.5 °C 375 to 1000 °C: ±0.004·|t| | −40 to 333 °C: ±2.5 °C 333 to 1200 °C: ±0.0075·|t| | −167 to 40 °C: ±2.5 °C −200 to −167 °C: ±0.015·|t| |
| R and S | 0 to 1100 °C: ±1 °C 1100 to 1600 °C: ±[1 + 0.003·(t − 1100)] | 0 to 600 °C: ±1.5 °C 600 to 1600 °C: ±0.0025·|t| | – |
| B | – | 600 to 1700 °C: ±0.0025·|t| | 600 to 800 °C: ±4 °C 800 to 1700 °C: ±0.005·|t| |
More information: thermocouple guide (types, construction and installation), cable colours, Pt100 calculator and 4–20 mA transmitters.